BOQU TPG-3030 (version 2.0) online total phosphorus analyzer measuring 0 to 500 mg/L TP in ranges, with no sample pretreatment.
Accuracyno more than±10% or no more than±0.2mg/L
Maintenance periodonce a month, each about 30 min.
NextTNG-3020 Total Nitrogen Analyzers
OutputRS-232; RS485; 4~20mA three ways
Power supplyAC230±10%V, 50±10%Hz, 5A
Reagent for value-based managementLess than 3 yuan/samples.
Product overview
Overview. The BOQU TPG-3030 (version 2.0) is an online total phosphorus analyzer whose sampling riser goes directly into the water, without sample pretreatment.
Ranges and accuracy. It measures 0 to 500 mg/L TP in ranges (0-2 mg/L, 0.1-10 mg/L, 0.5-50 mg/L, 1-100 mg/L, 5-500 mg/L). Accuracy is no more than ±10% or ±0.2 mg/L, and repeatability no more than ±5% or ±0.2 mg/L.
Measuring cycle. The minimum measuring period is 30 min, with digestion time adjustable from 5 to 120 min according to the water samples. The sampling period is adjustable from 10 to 9999 min, with a whole-point measurement mode.
Standard method. Measurement follows the national standard GB11893-89, "Water quality – Determination of total phosphorus Ammonium molybdate spectrophotometric method".
Calibration and maintenance. Calibration is programmable at any interval from 1 to 99 days. Monthly maintenance takes about 30 min. An abnormal alarm is provided, and a power failure does not lose data.
Outputs and integration. RS-232, RS485 and 4-20 mA outputs allow transmitting measurements to a supervision system. Control relies on a Panasonic PLC, and the water and electrical parts are separated.
Power and environment. Power supply AC 230 V ±10%, 50 ±10% Hz, 5 A. Recommended ambient temperature is 5 to 28 °C, with humidity of 90% or less and no condensation. Dimensions: 1570 x 500 x 450 mm (H*W*D).
National Standard GB11893-89 “Water quality – Determination of total phosphorus Ammonium molybdate spectrophotometric method”.
Next
TNG-3020 Total Nitrogen Analyzers
Others
abnormal alarm and power failure will not lose data;
Output
RS-232; RS485; 4~20mA three ways
Power supply
AC230±10%V, 50±10%Hz, 5A
Previous
TPG-3030 BOQU InstrumentTotal Phosphorus Analyzer
Reagent for value-based management
Less than 3 yuan/samples.
Repeatability
no more than ±5% or no more than ±0.2 mg/L
Sampling period
the time interval (10 ~ 9999min adjustable) and the whole point of measurement mode.
Size
1570 x500 x450mm(H*W*D).
Applications and use cases
Architectures in which TPG-3030 can play a role, derived from its category. This is not a compatibility statement: check the specifications against your requirements.
Answers to the questions that the recorded characteristics of TPG-3030 can answer. A question is not asked when nothing is recorded to answer it.
1Does it support RS-485?
Yes — recorded under "Output": RS-232; RS485; 4~20mA three ways.
2Does it support RS-232?
Yes — recorded under "Output": RS-232; RS485; 4~20mA three ways.
3Does it support CAN bus?
Yes — recorded under "Measurement Period": minimum measuring period of 30 min, according to actual water samples, can be modified at 5 ~ 120min arbitrary digestion time..
Similar parts to answer your needs. Compare the main characteristics. A recorded replacement is a documented relation; a similar product is a similarity only, never a compatibility claim.
Current productBOQU
TPG-3030
Online total phosphorus analyzer with multiple ranges
What is recorded about the regulatory status and the life of this part.
Regulatory compliance
No compliance statement is recorded for this part. Ask the supplier for the RoHS and REACH declarations and, for a radio product, the EU declaration of conformity.
No advertising or analytics cookies. A single, strictly necessary cookie is set only if you sign in to your account. Your quote basket and comparison stay in your browser. Details