TPG-3030 – Online total phosphorus analyzer with multiple ranges

Online total phosphorus analyzer with multiple ranges

General Industrial Sensors & Transmitters

Manufacturer part number
TPG-3030
Manufacturer
BOQU
Power supply
AC230±10%V, 50±10%Hz, 5A

BOQU TPG-3030 (version 2.0) online total phosphorus analyzer measuring 0 to 500 mg/L TP in ranges, with no sample pretreatment.

Product overview

Overview. The BOQU TPG-3030 (version 2.0) is an online total phosphorus analyzer whose sampling riser goes directly into the water, without sample pretreatment.

Ranges and accuracy. It measures 0 to 500 mg/L TP in ranges (0-2 mg/L, 0.1-10 mg/L, 0.5-50 mg/L, 1-100 mg/L, 5-500 mg/L). Accuracy is no more than ±10% or ±0.2 mg/L, and repeatability no more than ±5% or ±0.2 mg/L.

Measuring cycle. The minimum measuring period is 30 min, with digestion time adjustable from 5 to 120 min according to the water samples. The sampling period is adjustable from 10 to 9999 min, with a whole-point measurement mode.

Standard method. Measurement follows the national standard GB11893-89, "Water quality – Determination of total phosphorus Ammonium molybdate spectrophotometric method".

Calibration and maintenance. Calibration is programmable at any interval from 1 to 99 days. Monthly maintenance takes about 30 min. An abnormal alarm is provided, and a power failure does not lose data.

Outputs and integration. RS-232, RS485 and 4-20 mA outputs allow transmitting measurements to a supervision system. Control relies on a Panasonic PLC, and the water and electrical parts are separated.

Power and environment. Power supply AC 230 V ±10%, 50 ±10% Hz, 5 A. Recommended ambient temperature is 5 to 28 °C, with humidity of 90% or less and no condensation. Dimensions: 1570 x 500 x 450 mm (H*W*D).

See all specifications

Highlights

  • Accuracy: no more than±10% or no more than±0.2mg/L
  • Calibration period: 1~99 days, any interval, any time adjustable.
  • Maintenance period: once a month, each about 30 min.
  • Measuring Range: 0-500mg/L TP (0-2mg/L;0.1-10mg/L;0.5-50mg/L; 1-100mg/L;5-500mg/L)
  • Next: TNG-3020 Total Nitrogen Analyzers
  • Others: abnormal alarm and power failure will not lose data;

Typical applications

Derived from the part category, not a compatibility statement.

See also

Technical questions

Ask a question

No published question about this part yet. Questions are reviewed before publication.

Request a quote