Product overview
Overview. The Hollysys AW2000 integrated temperature transmitter converts universal input signals into a standardized 4 to 20 mA analog output. It accepts Resistance Temperature Detector (RTD) and Thermocouple (TC) inputs.
Main characteristics. The transmitter measures across a range of -200 to +1,600 ℃, with automatic cold-end temperature compensation and a nonlinear correction circuit. Its precision-grade conversion provides long-term stability and high measurement accuracy.
Interfaces and connectivity. The two-wire 4 to 20 mA output offers strong anti-interference performance. The 4 to 20 mA + HART version adds HART communication, with a high alarm at 21 mA and a low alarm at 3.8 mA.
Power and environment. Power supply is 12 to 40 VDC for the 4 to 20 mA version and 17 to 40 VDC for the 4 to 20 mA + HART version. Maximum load resistance follows RL = (U power supply - 12) / 0.024. Initialization takes at most 30 s at initial power-up and at most 6 s at normal power-up. Electrical isolation is 500 V between input and output, with a dielectric strength of 500 VAC for 1 min and an insulation resistance of 500 VDC for 30 s. The unit resists pressure, vibration and shock.
Typical uses. The transmitter is used to measure and transmit process temperature in the petrochemical, chemical, metallurgy, and food and pharmaceutical industries.
Integration. The 4 to 20 mA output allows the transmitter to be connected to a standard control or data acquisition system.