TF1100-EC – Clamp-on transit-time ultrasonic flow meter with LCD display

Clamp-on transit-time ultrasonic flow meter with LCD display

Ultrasonic Flow Meters

Manufacturer part number
20MM-5000MM
Manufacturer
Lanry

Wall-mounted clamp-on transit-time ultrasonic flow meter that measures the flow of liquids and liquefied gases in full pipes without cutting the pipe.

Product overview

Overview. The Lanry TF1100-EC is a wall-mounted transit-time ultrasonic flow meter. Its clamp-on transducers mount on the external surface of the pipe for non-invasive, non-intrusive flow measurement of liquids and liquefied gases in fully filled pipes. Three pairs of transducers cover the most common pipe diameter ranges.

Main characteristics. The measurement principle is the ultrasonic transit-time difference correlation. The flow velocity range is 0.01 to 12 m/s, bi-directional. Accuracy is ±1.0% of reading at rates above 0.3 m/s and ±0.003 m/s of reading below, with repeatability of 0.2% of reading. The display is a 4 line × 16 character backlit graphic LCD showing rate and velocity, with user-selectable damping of 0 to 99 s. Units are user configured in English or Metric.

Interfaces and connectivity. The transmitter provides a 4~20 mA output (accuracy 0.1%), a pulse output, a relay, an RS-232 port and an RS-485 port with Modbus support. A data logger function is built in. Totalizers cover gallons, ft³, barrels, lbs, liters, m³ and kg. Thermal energy measurement in GJ or KWh is optional, with a Pt1000 temperature sensor of ±0.1% accuracy.

Power and environment. The transmitter runs on 85-265 V AC or 24 V DC / 500 mA. Its fiberglass wall-mounted enclosure has an IP66 rating according to EN60529. The operating temperature range is -20℃ to +60℃, with a minimum operating temperature of -30 °C. Aluminum transducers suit standard liquid temperatures of -35℃ to 85℃, up to 120℃ for short periods, and PEEK transducers handle high temperatures.

Typical uses. The meter suits service and maintenance work, temporary replacement or verification of permanently installed meters, and performance measurement of pumps and regulating valves. Applications include water and wastewater, petrochemical and chemical processes, refrigeration and air conditioning, food and beverage, and pipeline leak detection. Supported liquids are clean or somewhat dirty, with turbidity below 10000 ppm.

Integration. The model connects to PLCs through its RS-485 Modbus port. Type S, M and L transducers cover pipe diameters of 20 to 50 mm, 40 to 1000 mm and 1000 to 6000 mm respectively.

Technical questions

Ask a question

No published question about this part yet. Questions are reviewed before publication.