TF1100-EC – Wall-mounted transit-time ultrasonic flow meter, clamp-on

Wall-mounted transit-time ultrasonic flow meter, clamp-on

Ultrasonic Flow Meters

  • RS-485
  • Modbus
  • 4-20 mA current loop
Manufacturer part number
DN20
Manufacturer
Lanry

Clamp-on ultrasonic flow meter for non-invasive measurement of liquid flow in full pipes, with 4-20 mA and RS485 Modbus outputs.

Product overview

Overview. The Lanry DN20, referenced TF1100-EC, is a wall-mounted transit-time ultrasonic flow meter whose transducers are mounted on the external surface of the pipe to measure the flow of liquids and liquefied gases in full pipes, without cutting the pipe or interrupting the process.

Main characteristics. Flow velocity range of 0.01 to 12 m/s, bi-directional. Accuracy of ±1.0% of reading above 0.3 m/s and ±0.003 m/s of reading below 0.3 m/s, repeatability of 0.2% of reading, response time of 0.5 s, displayed value damping adjustable from 0 to 99 s. Backlit 4 line × 16 character graphic LCD, units user configurable in English and Metric.

Interfaces and connectivity. 4-20 mA output (accuracy 0.1%), relay, RS232 and RS485 with Modbus, data logger function. Keypad lockout and system lockout for access security.

Power and environment. AC 85-265 V or DC 24 V / 500 mA supply. Degree of protection IP66 according to EN60529, operating temperature -20 °C to +60 °C, minimum operating temperature -30 °C. Fiberglass housing, size 244 × 196 × 114 mm, weight 2.4 kg.

Transducers. Aluminum transducers for standard temperature and PEEK for high temperature, covering pipe diameters of 20 to 50 mm (type S), 40 to 1000 mm (type M) and 1000 to 6000 mm (type L). Integrated Pt1000 temperature sensor, accuracy ±0.1%. Clean or somewhat dirty liquids with turbidity below 10000 ppm. Optional thermal energy measurement in GJ or kWh.

Typical uses. Service and maintenance, replacement of defective meters, control of pumps and regulating valves, water and wastewater, hot and cooling water circuits, chemical and petrochemical industries, food, beverage and pharmaceutical industries.

Integration. RS485 connection with Modbus allows integration into a PLC or a supervision system.