TF1100-EC – Wall-mount clamp-on transit-time ultrasonic flow meter

Wall-mount clamp-on transit-time ultrasonic flow meter

Ultrasonic Flow Meters

  • -20℃ to +60℃
  • 0.5s
  • Std:10m
  • Wall-mounted
Manufacturer part number
DN20-DN6000
Manufacturer
Lanry
Operating Temperature
-20℃ to +60℃
Response time
0.5s

Clamp-on transit-time ultrasonic flow meter for measuring liquid flow in a fully filled pipe without cutting the pipe. Used for maintenance, control and temporary replacement of permanently installed meters.

Product overview

Overview. The Lanry TF1100-EC is a wall-mount transit-time ultrasonic flow meter. Its clamp-on transducers are mounted on the external surface of the pipe for non-invasive and non-intrusive flow measurement of liquids and liquefied gases in fully filled pipes.

Main characteristics. Flow velocity range of 0.01 to 12 m/s, bi-directional. Accuracy of ±1.0% of reading above 0.3 m/s and ±0.003 m/s below. Repeatability of 0.2% of reading. Pipe diameters from 20 to 6000 mm depending on transducer type S, M or L. Flow rate and velocity display, totalisation in gallons, ft³, barrels, lbs, liters, m³ and kg, units user-configurable in English and Metric. Data logger function and optional thermal energy measurement in GJ or kWh.

Interfaces and connectivity. 4~20 mA, relay, RS232 and RS485 outputs with Modbus protocol. Temperature measured by a clamp-on Pt1000 sensor, accuracy ±0.1%.

Power and environment. Power supply 85-265 V AC or 24 V DC, 500 mA. Fiberglass housing, IP66 degree of protection according to EN60529. Operating temperature -20 °C to +60 °C, liquid temperature -35 °C to 85 °C for short periods up to 120 °C. Dimensions 244 × 196 × 114 mm, weight 2.4 kg. Backlit 4-line × 16-character graphic LCD display, keypad lockout.

Typical uses. Service and maintenance, performance checks of pumps and regulating valves, temporary replacement of permanently installed meters, in water networks, chemical and petrochemical industry, air conditioning and power generation.

Integration. The RS485 signal with Modbus protocol allows connection to a programmable logic controller or a supervision system.

See all specifications

Highlights

  • Response time: 0.5s
  • Cable Length: Std:10m
  • Communication: 4~20mA(accuracy 0.1%),OCT, Relay, RS232, RS485 (Modbus),data logger
  • Damping of displayed value: 0-99s(selectable by user)
  • Degree of protection: IP66 according to EN60529
  • Display: 4 line×16 English letters LCD graphic display, backlit

Good to know before integrating

  • Thermal: rated operating range -20℃ to +60℃ — check it against the installation environment.

Technical questions

Ask a question

No published question about this part yet. Questions are reviewed before publication.