TF1100-CH – Handheld transit-time clamp-on ultrasonic flow meter

Handheld transit-time clamp-on ultrasonic flow meter

Ultrasonic Flow Meters

Manufacturer part number
TF1100-CH
Manufacturer
Lanry

Handheld transit-time ultrasonic flow meter measuring the flow of liquids in fully filled pipes without intrusion, using sensors mounted on the outside of the pipe. Suited to maintenance and meter checks.

Product overview

Lanry TF1100-CH. this handheld transit-time ultrasonic flow meter measures the flow of liquids and liquefied gases in fully filled pipes, using sensors mounted on the external surface of the pipe.

Measurement principle. measurement relies on the ultrasonic transit-time difference correlation principle. The transducers are non-invasive and do not contact the fluid. Three pairs of transducers cover the most common pipe diameter ranges, from 20 to 50 mm for types S or B and from 40 to 5000 mm for types M or A.

Performance. flow velocity range of 0.01 to 12 m/s, bi-directional. Accuracy of ±1.0% of reading at rates above 0.3 m/s and ±0.003 m/s of reading at rates below 0.3 m/s. Repeatability of 0.2% of reading, resolution of 0.25 mm/s and response time of 0.5 s. Damping of the displayed value is adjustable from 0 to 99 s by the user.

Display and data. backlit 4-line by 16-character LCD graphic display. Rate and velocity displayed, totalized in gallons, ft³, barrels, lbs, liters, m³ and kg, with user-configured units in English or metric. Data logger function included.

Power and environment. AC power supply 85 to 265 V, up to 14 hours on fully charged internal batteries. Degree of protection IP65 according to EN60529. Operating temperature from -20 °C to +60 °C, minimum operating temperature -30 °C. ABS housing (UL 94HB), 212 × 100 × 36 mm, 0.5 kg, handheld enclosure.

Transducers. standard 5 m cable. Aluminum transducers for standard temperature and PEEK for high temperature, for liquids from -35 °C to 85 °C in the standard version. Works with clean and somewhat dirty liquids with turbidity below 10000 ppm.

Typical uses. suited to service and maintenance, checking meters in place, flow rate calibration, data comparison and checking the running status of meters. It can also be used for the temporary replacement of a permanently installed meter. Security is provided by keypad and system lockout.

Connectivity. RS-232 port for transmitting logged data.

Available on quotation from Atlas Electronics.

Technical questions

Ask a question

No published question about this part yet. Questions are reviewed before publication.