TF1100-DC – Dual-channel transit-time ultrasonic flow meter, wall-mounted

Dual-channel transit-time ultrasonic flow meter, wall-mounted

Ultrasonic Flow Meters

  • -10℃ to +60℃
  • 0.5s
  • Std:10m
  • Wall-mounted
Manufacturer part number
ULTRASONIC-FLOW-METER-HIGH-QUALITY
Manufacturer
Lanry
Operating Temperature
-10℃ to +60℃
Response time
0.5s

Clamp-on transit-time ultrasonic flow meter for non-invasive flow measurement of liquids in fully filled pipes, with dual channel and RS485 Modbus output.

Product overview

Overview. The Lanry TF1100-DC is a wall-mounted transit-time ultrasonic flow meter working on the transit-time difference correlation principle. Clamp-on transducers mount on the external surface of the pipe for non-invasive and non-intrusive measurement of liquid and liquefied gas flow in fully filled pipes.

Main characteristics. Accuracy of ±0.5% of reading at rates above 0.3 m/s, repeatability of 0.15% of reading, bi-directional flow velocity range of 0.01 to 15 m/s, resolution of 0.1 mm/s and response time of 0.5 s. Pipe diameter range is 20-50 mm for type B and 40-4000 mm for type A. Transducers are available in aluminum or stainless steel SUS304.

Interfaces and connectivity. 4~20 mA output (accuracy 0.1%), OCT, relay, RS485 with Modbus protocol and data logger. 4.3-inch colour LCD with 5-line display and 16 keys. Thermal energy measurement optional, in GJ or kWh, with Pt1000 temperature sensor from 0 to 200 °C.

Power and environment. Power supply AC 85-265 V or DC 24 V/500 mA. Degree of protection IP66 according to EN60529, operating temperature -10 °C to +60 °C, suited liquid temperature -35 °C to 200 °C. Fiberglass housing, 244 × 196 × 114 mm, 2.4 kg.

Typical uses. Service and maintenance, replacement of defective meters, monitoring of pumps and regulating valves, measurement of hot water, cooling water, potable water and sea water, and of chemical liquids with turbidity below 10000 ppm.

Integration. RS485 Modbus output allows connection to a PLC or supervision system, and the totalizer covers common volume and mass units.

See all specifications

Highlights

  • Response time: 0.5s
  • Cable Length: Std:10m
  • Communication: 4~20mA(accuracy 0.1%),OCT, Relay, RS485 (Modbus),data logger
  • Damping of displayed value: 0-99s(selectable by user)
  • Degree of protection: IP66 according to EN60529
  • Display: 4.3’’ color LCD 5 lines display, 16 keys

Good to know before integrating

  • Thermal: rated operating range -10℃ to +60℃ — check it against the installation environment.