WQ-CLI-2A
Chloride ion sensor RS485 for industrial water

- Operating temperature and pressure
- -5 to +65 degrees C; pressure <0.2 MPa
- Medium pH
- 2-12
- Products
- Weather Stations
Chloride ion sensor RS485 for industrial water
Ion-selective chloride sensor for monitoring chloride in water, with RS485 Modbus RTU output, intended for industrial water.
Overview. The Niubol NBL-WQ-CLI-2A is an ion-selective electrode chloride sensor that measures chloride concentration in water and transmits the value over RS485 Modbus RTU. It is not a residual chlorine instrument.
Main characteristics. Range of 0-3500 mg/L at 0.1 mg/L resolution, or 0-35000 mg/L at 1 mg/L resolution. Accuracy of ±5 %, temperature ±0.3 °C. Pt1000 automatic temperature compensation and two-point calibration. Medium pH range of 2 to 12.
Connectivity and power. Supply of 12 to 24 V DC, consumption of 0.2 W at 12 V. RS485 Modbus RTU output for connection to a PLC or supervisory system. Standard cable of 5 m; alternative lengths and jacket requirements to be specified.
Housing and installation. POM body, immersion mounting with 3/4 NPT thread, IP68 protection rating. Operating temperature from -5 to +65 °C, pressure below 0.2 MPa.
Typical uses. Following chloride changes during salt intrusion, industrial water mixing or on a defined treatment stream, with readings transmitted to a Modbus system.
Points to check. Sulfide, iodide and bromide can interfere with the measurement; assess their presence in the actual stream before selecting the range, and compare results against a reference method on the installed matrix.
No published question about this part yet. Questions are reviewed before publication.
Every characteristic recorded for WQ-CLI-2A, with where it comes from. Use the filters to find one quickly.
| Characteristic | Value |
|---|---|
| Accuracy | +/-5%; temperature +/-0.3 degrees C; confirm percentage basis and applicable matrix |
| Cable | Standard 5 m; alternative length and jacket requirements to be specified |
| Compensation and calibration | Pt1000 automatic temperature compensation; two-point calibration |
| Housing and installation | POM; immersion mounting with 3/4 NPT thread; IP68 |
| Medium pH | 2-12 |
| Operating temperature and pressure | -5 to +65 degrees C; pressure <0.2 MPa |
| Products | Weather Stations |
| Range options and resolution | 0-3500 mg/L at 0.1 mg/L resolution; 0-35000 mg/L at 1 mg/L resolution |
| Supply, output and consumption | 12-24 V DC; RS485 Modbus RTU; 0.2 W at 12 V |
No specification matches.
Last verification : 2026-11-02
Official documents and the technical data recorded for WQ-CLI-2A. For a missing document or a specific revision, our experts send it to you.
Ion-selective chloride sensor for monitoring chloride in water, with RS485 Modbus RTU output, intended for industrial water.
| Parameter | Value |
|---|---|
| Accuracy | +/-5%; temperature +/-0.3 degrees C; confirm percentage basis and applicable matrix |
| Cable | Standard 5 m; alternative length and jacket requirements to be specified |
| Compensation and calibration | Pt1000 automatic temperature compensation; two-point calibration |
| Housing and installation | POM; immersion mounting with 3/4 NPT thread; IP68 |
| Medium pH | 2-12 |
| Operating temperature and pressure | -5 to +65 degrees C; pressure <0.2 MPa |
| Products | Weather Stations |
| Range options and resolution | 0-3500 mg/L at 0.1 mg/L resolution; 0-35000 mg/L at 1 mg/L resolution |
| Supply, output and consumption | 12-24 V DC; RS485 Modbus RTU; 0.2 W at 12 V |
Points to check before using WQ-CLI-2A in a design. They are derived from the recorded characteristics, not from the manufacturer's application circuits.
| Supply, output and consumption | 12-24 V DC; RS485 Modbus RTU; 0.2 W at 12 V |
Check the exact pin assignment and the application circuit in the datasheet, or ask our experts.
Answers to the questions that the recorded characteristics of WQ-CLI-2A can answer. A question is not asked when nothing is recorded to answer it.
Yes — recorded under "Supply, output and consumption": 12-24 V DC; RS485 Modbus RTU; 0.2 W at 12 V.
Yes — recorded under "Supply, output and consumption": 12-24 V DC; RS485 Modbus RTU; 0.2 W at 12 V.
-5 to +65 degrees C; pressure <0.2 MPa (source: Operating temperature and pressure).
Similar parts to answer your needs. Compare the main characteristics. A recorded replacement is a documented relation; a similar product is a similarity only, never a compatibility claim.
Chloride ion sensor RS485 for industrial water

MPP1000 Multi-parameter analyzer

MDX500 Multi-parameter analyzer

SUP-ADP2400 Digital pH Sensor for Reliable Online Water Quality Monitoring

Describe your application and constraints in a request.
What is recorded about the regulatory status and the life of this part.
No compliance statement is recorded for this part. Ask the supplier for the RoHS and REACH declarations and, for a radio product, the EU declaration of conformity.
| Current status | Status not stated |
| Key dates | Not recorded |
| Manufacturer part number | WQ-CLI-2A |
| Manufacturer | Niubol |
pH and Conductivity Analyzers
RS485 Modbus Dissolved Oxygen Sensor, Fluorescence Method
Price on request
pH and Conductivity Analyzers
Online UV fluorescence BOD sensor, reagent-free · Ref. : WQ-BOD-4S
Price on request