NYCSUL-N – 24 VDC ultrasonic mud-water interface level transmitter

24 VDC ultrasonic mud-water interface level transmitter

Level Transmitters

  • 0~5m,0~10m
  • 5m,10m
  • PG11
  • G3/4 or G2
Manufacturer part number
32487714-00
Manufacturer
Nuoying
Measuring Range
0~5m,0~10m
Range
5m,10m

Non-contact ultrasonic level transmitter for mud-water interface measurement, with 0 to 5 m and 0 to 10 m ranges, a 4-20 mA analog output and an RS485 Modbus-RTU digital output.

Product overview

Overview. The Nuoying NYCSUL-N is an ultrasonic mud-water interface level transmitter made by Nuoying, designed for non-contact measurement of the mud-water interface. It is supplied in a split, four-wire version.

Main characteristics. Measuring ranges of 0 to 5 m and 0 to 10 m, resolution of 1 mm, accuracy of ±1 %FS under standard conditions and a beam angle of 12 degrees. The dead band is less than 0.5 m or 0.8 m and the required depth of draught is at least 0.5 m. The display shows mud position, distance, current, echo waveform and historical curve, in Chinese and English.

Interfaces and communication. 4-20 mA analog output (load less than 500 ohms), RS485 digital output with Modbus-RTU protocol (HART 5.0 selectable), and two switch outputs with contact rating 3 A 250 VAC / 5 A 30 VDC. Electrical interface PG11 and process connection G3/4 or G2. The probe is made of 304 stainless steel and the instrument housing of ABS.

Power and environment. Power supply DC 12 V to 36 V / 30 mA or AC 85 V to 265 V / 5 W. Ambient temperature −20 °C to +60 °C and process temperature 0 °C to +60 °C. Waterproof grade IP65 for the instrument and IP68 for the probe. Probe cable of 10 m, extendable to 60 m.

Typical uses and integration. Thanks to its RS485 Modbus-RTU output, the device can be connected to a PLC or a supervision system. It carries the CE certification. Parameters can be set on site with the built-in keys or remotely via RS485. Available on quotation from Atlas Electronics.

See all specifications

Highlights

  • Accuracy: ±1%FS* (Standard condition *)
  • Ambient temperature: -20℃~+60℃*
  • Analog Output: 4~20mA/bit12 current output, load less than 500 ohms
  • Beam Angle: Angle 12 degrees
  • Certification: CE
  • Dead band: Less than 0.5m,Less than 0.8m

See also

Technical questions

Ask a question

No published question about this part yet. Questions are reviewed before publication.