OCH2B10 – 2-megapixel 1500 x 1500 CMOS image sensor for endoscopes

2-megapixel 1500 x 1500 CMOS image sensor for endoscopes

Image Sensors

  • 650 µA
  • image array
Manufacturer part number
OCH2B10
Manufacturer
OmniVision
Power Requirement Standby
650 µA
Standby
650 µA

Square 2-megapixel CMOS image sensor in 1/7.5" format, housed in a 2.5 x 2.5 mm CameraCubeChip® package for disposable and reusable endoscopes.

Product overview

Overview. The OmniVision OCH2B10 is a 2-megapixel (1500 x 1500) CMOS image sensor in 1/7.5" format, integrated in a 2.5 x 2.5 mm CameraCubeChip® module designed for endoscopes. It is a rolling-shutter CMOS matrix sensor with a CFA colour filter, built on PureCel®Plus-S technology.

Main characteristics. The sensor delivers 60 frames per second at full resolution, with a 1.116 µm pixel size and an image area of 1691.856 x 1691.856 µm. It supports 1500 x 1500, 960 x 540, 1280 x 720 and 640 x 480 image sizes, as well as horizontal and vertical subsampling. The chip includes automatic black level calibration (ABLC), dynamic DPC support, a built-in temperature sensor and two on-chip phase lock loops (PLLs).

Interfaces and connectivity. The digital output uses the 4-wire AntLinx® interface at speeds up to 1.6 Gbps, which supports a 4-metre connection between the endoscope camera and the camera control unit. Configuration is done through the standard serial SCCB interface. Output formats include 10-bit RGB non-HDR, and Stagger HDR 2 exposure timing is supported.

Controls and settings. Programmable functions include frame rate, mirror and flip, binning, cropping and windowing.

Power and environment. Standby power consumption is 650 µA. The module is biocompatible and waterproof, and is not autoclavable. OmniVision is ISO 13485 certified, and medical products are tested for EMC and EMI.

Typical uses. The sensor targets endoscopes for ear, nose and throat and airway applications (laryngoscope, bronchoscope, otoscope, rhinoscope), brain, eye, heart and spine applications (neuroendoscope, ophthalmoscope), arthroscopy, stomach and intestine applications (duodenoscope, proctoscope) and intra-oral scanners.

See all specifications

Highlights

  • Power Requirement Standby: 650 µA
  • Standby: 650 µA
  • tceles wor: image array