| Applicable Fiber Type | SM (9.5/125um), GI (50/125um, 62.5 um, 125um), Large core diameter fiber (maximum 200um) |
| Dimension | 426 (W) × 221 (H) × 450 (D) mm |
| Display | 12.1 inch touch screen |
| Dynamic Range | • Resolution 0.02nm: 60dB (peak ±0.2nm), 46dB (peak ±0.1nm); • Resolution 0.05nm: 73dB (peak ±1.0nm), 63dB (peak ±0.4nm), 52dB (peak ±0.2nm); • Resolution 0.1nm: 60dB (peak ±0.4nm), 46dB (peak ±0.2nm) |
| Instrument Interface | USB/Ethernet/GPIB/RS232C/VGA |
| Maximum Input Power | +20 dBm (per channe, full wavelength band) |
| Maximum Safe Input Power | +25 dBm (total input power) |
| Optical Output Options | Standard C-band DFB calibration light source; DFB/FP light source (standard 1550nm, other wavelengths are optional); SLED light source (band range optional),; SLED+C2H2 light source; Other light source type accessories can be customized |
| Optical Return Loss | 35dB (using APC connectors) |
| Polarization Dependence | ±0.05 dB (1550nm), ±0.08 dB (1310nm) |
| Power Accuracy | ±0.4 dB (1310/1550nm, input power: -20dBm, sensitivity: MID) |
| Power Flatness | • ±0.1dB (1520nm to 1580nm); • ±0.2dB (1450nm to 1520nm, 1580nm to 1620nm) |
| Power Linearity | ±0.05 dB (input power: -50dBm to +10dBm) |
| Power Sensitivity | • -90 dBm (1300 to 1620nm); • -85 dBm (1000 to 1300nm); • -60 dBm (600 to 1000nm) (sensitivity: HIGH3) |
| Power Sensitivity Setting | NORMAL, MID, HIGH1, HIGH2 and HIGH3 |
| S6362D-H02 — Built-in light source option | Light source type: SLED Power range: ≥1mW Center wavelength: 1550nm |
| S6362D-H03 — Build-in Source | Type: SLED + C2H2 absorption tank for higher calibration, Power: ≥1mW, Center Wavelength: 1550nm |
| S6362D-H04 — Interface options | Contains GPIB interface, trigger input interface, trigger output interface |
| S6362D — Optical Spectrum Analyzer | 600 nm to 1700 nm |
| S6362D-S02 — High Accuracy Analysis Software | Optimized wavelength measurement accuracy and dynamic range. Wavelength accuracy: ±10pm (1520nm to 1580nm), Dynamic range: better than 75dB (Typ: 76dB) |
| Sampling Points | 101 to 50001, AUTO |
| Sampling Resolution | Minimum 0.001nm |
| Scan Span | 0.2 nm to 1100 nm (full span), 0 nm |
| Spectral range | 600nm – 1700nm |
| Stray Light Suppression Rate | 76dB |
| Wavelength Accuracy | • ±0.02 nm (1520nm to 1620nm); • ±0.04 nm (1450nm to 1520nm); • ±0.10 nm (full span) |
| Wavelength Linearity | ±0.01nm (1520 – 1580nm) |
| Wavelength Repeatability | ±0.005nm (2 minutes) |
| Wavelength Resolution Setting | 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2 nm |
| Weight | 19kg |