STB8817B
Benchtop LCR meter, 50 Hz to 100 kHz

- Basic Accuracy
- 0.1%
- AC Test Signal Level
- 0.01V-1V
- V/I Monitor
- Yes
Benchtop LCR meter, 50 Hz to 100 kHz
The Saluki Technology STB8817B is a benchtop LCR meter measuring L, C, R, |Z|, D and Q on passive components and impedance elements, with 0.1% basic accuracy.
Overview. The Saluki Technology STB8817B is a benchtop LCR meter, an impedance measuring instrument that automatically determines inductance L, capacitance C, resistance R, impedance modulus |Z|, quality factor Q and loss tangent D of components.
Main characteristics. Basic accuracy is 0.1%. The AC test signal level ranges from 0.01 V to 1 V. Measurement offers ten test frequencies from 50 Hz to 100 kHz (50 Hz, 60 Hz, 100 Hz, 120 Hz, 1 kHz, 10 kHz, 20 kHz, 40 kHz, 50 kHz, 100 kHz). A ten-point list sweep allows each scan point to be sorted individually. Measured parameters include L, C, R, |Z|, D, Q, X, θd, θr, Vm, Im and △%, with V/I monitoring.
Accessories. The fixture range includes magnetic ring test fixtures, SMD component test fixtures, four-terminal test fixtures, a dielectric test fixture and a fixture for DC_LINK capacitors. The fixtures accept a maximum DC bias of ±42 V. The instrument is supplied with a short circuit, a fixture and four-terminal Kelvin test leads.
Typical uses. It suits the measurement of passive components such as capacitors, inductors, magnetic cores, resistors and transformers, and the impedance evaluation of printed circuit boards, relays, cables and batteries, for production inspection as well as research and development.
No published question about this part yet. Questions are reviewed before publication.
Every characteristic recorded for STB8817B, with where it comes from. Use the filters to find one quickly.
| Characteristic | Value |
|---|---|
| AC Test Signal Level | 0.01V-1V |
| Basic Accuracy | 0.1% |
| Frequency Range | 50Hz, 60Hz, 100Hz, 120Hz, 1kHz, 10kHz, 20kHz, 40kHz, 50kHz, 100kHz |
| List Sweep | 10-point list sweep, each scan point can be individually sorted |
| SBF0007A — Magnetic Ring Test Fixture | DC-120MHz, maximum ±42V DC bias, applied to single-turn inductance test of small magnetic ring, customizable size |
| SBF0008A — SMD Component Test Fixture | DC-120MHz, maximum ±42V DC bias, applied to SMD components, especially high-frequency small capacitor ≤3pF or small inductance ≤1uH, test frequency ≥100kHz |
| SBF0009B — SMD Component Test Fixture | DC-15MHz, maximum ±42V DC bias, applied to SMD components |
| SBF0047 — Four-terminal Test Fixture | DC-120MHz, maximum ±42V DC bias, applied to impedance testing of wire-like devices, with shielded ground terminal |
| SBF0062A — Four-terminal Test Fixture | DC-100kHz, maximum ±42V DC bias, applied to test thin film bulk DC_LINK capacitors for electric vehicles |
| SBF0063 — Four-terminal Test Fixture | DC-100kHz, maximum ±42V DC bias, applied to test bolted capacitor and DC_LINK capacitors |
| SBF0077 — Dielectric Test Fixture | DC-120MHz, DUT size: 10mm-56mm, DUT thickness: ≤10mm, applied to dielectric analysis of solid materials |
| SBF1108C — Four-terminal SMD Test Fixture | DC-40MHz, maximum ±42V DC bias, applied to SMD components, especially high-frequency small capacitor ≤3pF or small inductance ≤1uH, test frequency ≥100kHz |
| STB8817B — LCR Meter | Frequency range: 50Hz-100kHz, including short circuit, fixture and Kelvin test lead with box four-terminal insulation & lock |
| Test Parameters | L, C, R, |Z|, D, Q, X, θd, θr, Vm, Im, △% |
| V/I Monitor | Yes |
No specification matches.
Last verification : 2026-11-03
Official documents and the technical data recorded for STB8817B. For a missing document or a specific revision, our experts send it to you.
The Saluki Technology STB8817B is a benchtop LCR meter measuring L, C, R, |Z|, D and Q on passive components and impedance elements, with 0.1% basic accuracy.
| Parameter | Value |
|---|---|
| AC Test Signal Level | 0.01V-1V |
| Basic Accuracy | 0.1% |
| Frequency Range | 50Hz, 60Hz, 100Hz, 120Hz, 1kHz, 10kHz, 20kHz, 40kHz, 50kHz, 100kHz |
| List Sweep | 10-point list sweep, each scan point can be individually sorted |
| SBF0007A — Magnetic Ring Test Fixture | DC-120MHz, maximum ±42V DC bias, applied to single-turn inductance test of small magnetic ring, customizable size |
| SBF0008A — SMD Component Test Fixture | DC-120MHz, maximum ±42V DC bias, applied to SMD components, especially high-frequency small capacitor ≤3pF or small inductance ≤1uH, test frequency ≥100kHz |
| SBF0009B — SMD Component Test Fixture | DC-15MHz, maximum ±42V DC bias, applied to SMD components |
| SBF0047 — Four-terminal Test Fixture | DC-120MHz, maximum ±42V DC bias, applied to impedance testing of wire-like devices, with shielded ground terminal |
| SBF0062A — Four-terminal Test Fixture | DC-100kHz, maximum ±42V DC bias, applied to test thin film bulk DC_LINK capacitors for electric vehicles |
| SBF0063 — Four-terminal Test Fixture | DC-100kHz, maximum ±42V DC bias, applied to test bolted capacitor and DC_LINK capacitors |
| SBF0077 — Dielectric Test Fixture | DC-120MHz, DUT size: 10mm-56mm, DUT thickness: ≤10mm, applied to dielectric analysis of solid materials |
| SBF1108C — Four-terminal SMD Test Fixture | DC-40MHz, maximum ±42V DC bias, applied to SMD components, especially high-frequency small capacitor ≤3pF or small inductance ≤1uH, test frequency ≥100kHz |
| STB8817B — LCR Meter | Frequency range: 50Hz-100kHz, including short circuit, fixture and Kelvin test lead with box four-terminal insulation & lock |
| Test Parameters | L, C, R, |Z|, D, Q, X, θd, θr, Vm, Im, △% |
| V/I Monitor | Yes |
Answers to the questions that the recorded characteristics of STB8817B can answer. A question is not asked when nothing is recorded to answer it.
Yes — recorded under "List Sweep": 10-point list sweep, each scan point can be individually sorted.
Similar parts to answer your needs. Compare the main characteristics. A recorded replacement is a documented relation; a similar product is a similarity only, never a compatibility claim.
Benchtop LCR meter, 50 Hz to 100 kHz

Series Precision LCR Meter

LCR Meter

Describe your application and constraints in a request.
What is recorded about the regulatory status and the life of this part.
No compliance statement is recorded for this part. Ask the supplier for the RoHS and REACH declarations and, for a radio product, the EU declaration of conformity.
| Current status | Status not stated |
| Key dates | Not recorded |
| Manufacturer part number | STB8817B |
| Manufacturer | Saluki Technology |
No advertising or analytics cookies. A single, strictly necessary cookie is set only if you sign in to your account. Your quote basket and comparison stay in your browser. Details