KEW_6010B
Multi-function installation tester with earth and RCD testing

- Impedance range
- 20/2000&Omega
- Measuring Range
- 20/200Ω (Auto-Ranging)
- Rated Current
- > 1mA
- Short Circuit Current
- > 200mA
Multi-function installation tester with earth and RCD testing
Installation and Earth Testers
The Kyoritsu KEW 6010B is a multi-function installation tester compliant with IEC 61557, designed for earth resistance measurement and testing of residual current devices.
Kyoritsu KEW 6010B multifunction installation tester — the KEW 6010B is a multi-function installation tester designed to IEC 61010-1 and IEC 61557, covering earth resistance measurement and testing of residual current devices.
Main characteristics. Impedance measuring range of 20 and 2000 Ω with auto-ranging, test voltage of 500 V and 1000 V, open circuit voltage above 6 V and short circuit current above 200 mA. Accuracy of ±(3 % rdg + 3 dgt) on impedance and ±(1 % rdg + 3 dgt) on trip time. Data memory of 300 measured results.
Residual current device testing. Test currents of 5 mA for IΔn of 10 mA, 15 mA for 30 and 100 mA, 150 mA for 300 and 500 mA. Test modes x1/2, x1, DC, Fast and Auto Ramp, with accuracy of ±10 % in DC and ±4 % in Auto Ramp.
Power and environment. Powered by 8 R6/LR6 batteries, rated voltage 230 V +10 to 15 % (50 Hz). IP40 protection according to IEC 60529, dimensions 115 × 175 × 86 mm, weight approx. 840 g.
Accessories and communication. Test leads 7122B, test lead with IEC connector KAMP10, fuse 8923 0.5 A/600 V supplied with one spare, cord case 9092 and shoulder strap 9121. Results transfer to a PC through the optical RS-232C port using the optional 8212-USB adapter. Distribution board test leads 7133B available as an option.
Typical uses. Commissioning and inspection of low-voltage electrical installations, verification of continuity and earth resistance, testing of residual current circuit breakers. Available on quotation from Atlas Electronics.
No published question about this part yet. Questions are reviewed before publication.
Every characteristic recorded for KEW_6010B, with where it comes from. Use the filters to find one quickly.
| Characteristic | Value |
|---|---|
| Accessories | 7122B (Test leads); KAMP10 (Test lead with IEC connector); 8923 (Fuse[0.5A/600V] × 1 (included), 1 (spare); 9092 (Cord case); 9121 (Shoulder strap); Shoulder pad; Instruction manual, R6(AA) × 8 |
| Accuracy | ±(3%rdg + 3dgt) |
| Accuracy — Test Current | x1/2: -8 to 2%; x1, FAST: +2%+8%; DC: ±10%; Auto Ramp: ±4% |
| Accuracy — Trip Time | ±(1%rdg + 3dgt) |
| Applicable Standards | IEC 61010-1 CAT Ⅲ 300V Pollution Degree 2, IEC 61010-2-034, IEC 61557-1, 2, 3, 4, 6, 10, IEC 60529 (IP40) |
| Data Memory | 300 measured results |
| Dimension | 115(L) × 175(W) × 86(D)mm |
| Impedance range | 20/2000&Omega |
| Measuring Range | 20/200Ω (Auto-Ranging) |
| Normal Test Current | 20Ω :25A/10ms; 2000Ω :15mA/350ms max. |
| Open circuit voltage | > 6V |
| Optional Accessories | 7133B (Distribution board test leads); 8212-USB (USB adapter) |
| Power Source | R6/LR6 × 8 |
| Product Catalog | KEW 6010B Product Catalog (English)(8.03 MB) |
| Rated Current | > 1mA |
| Rated Voltage | 230V + 10 to 15%(50Hz) |
| Short Circuit Current | > 200mA |
| Test Current | 5mA at IΔ n = 10mA; 15mA at IΔ n = 30/100mA; 150mA at IΔ n = 300/500mA |
| Test Current; (Test Current Duration) — Auto Ramp | Goes up by 10% from 20% to 110% of IΔ n. 300ms×10 |
| Test Current; (Test Current Duration) — DC | 10, 30, 100, 300mA (2000ms), 500mA (200ms) |
| Test Current; (Test Current Duration) — Fast | 150mA (50ms) |
| Test Current; (Test Current Duration) — x1/2, x1 | 10, 30, 100, 300, 500mA (2000ms) |
| Test voltage | 500/1000V |
| Weight | Approx. 840g |
No specification matches.
Last verification : 2026-11-03
Official documents and the technical data recorded for KEW_6010B. For a missing document or a specific revision, our experts send it to you.
The Kyoritsu KEW 6010B is a multi-function installation tester compliant with IEC 61557, designed for earth resistance measurement and testing of residual current devices.
| Parameter | Value |
|---|---|
| Accessories | 7122B (Test leads); KAMP10 (Test lead with IEC connector); 8923 (Fuse[0.5A/600V] × 1 (included), 1 (spare); 9092 (Cord case); 9121 (Shoulder strap); Shoulder pad; Instruction manual, R6(AA) × 8 |
| Accuracy | ±(3%rdg + 3dgt) |
| Accuracy — Test Current | x1/2: -8 to 2%; x1, FAST: +2%+8%; DC: ±10%; Auto Ramp: ±4% |
| Accuracy — Trip Time | ±(1%rdg + 3dgt) |
| Applicable Standards | IEC 61010-1 CAT Ⅲ 300V Pollution Degree 2, IEC 61010-2-034, IEC 61557-1, 2, 3, 4, 6, 10, IEC 60529 (IP40) |
| Data Memory | 300 measured results |
| Dimension | 115(L) × 175(W) × 86(D)mm |
| Impedance range | 20/2000&Omega |
| Measuring Range | 20/200Ω (Auto-Ranging) |
| Normal Test Current | 20Ω :25A/10ms; 2000Ω :15mA/350ms max. |
| Open circuit voltage | > 6V |
| Optional Accessories | 7133B (Distribution board test leads); 8212-USB (USB adapter) |
| Power Source | R6/LR6 × 8 |
| Product Catalog | KEW 6010B Product Catalog (English)(8.03 MB) |
| Rated Current | > 1mA |
| Rated Voltage | 230V + 10 to 15%(50Hz) |
| Short Circuit Current | > 200mA |
| Test Current | 5mA at IΔ n = 10mA; 15mA at IΔ n = 30/100mA; 150mA at IΔ n = 300/500mA |
| Test Current; (Test Current Duration) — Auto Ramp | Goes up by 10% from 20% to 110% of IΔ n. 300ms×10 |
| Test Current; (Test Current Duration) — DC | 10, 30, 100, 300mA (2000ms), 500mA (200ms) |
| Test Current; (Test Current Duration) — Fast | 150mA (50ms) |
| Test Current; (Test Current Duration) — x1/2, x1 | 10, 30, 100, 300, 500mA (2000ms) |
| Test voltage | 500/1000V |
| Weight | Approx. 840g |
Answers to the questions that the recorded characteristics of KEW_6010B can answer. A question is not asked when nothing is recorded to answer it.
Yes — recorded under "Optional Accessories": 7133B (Distribution board test leads); 8212-USB (USB adapter).
Similar parts to answer your needs. Compare the main characteristics. A recorded replacement is a documented relation; a similar product is a similarity only, never a compatibility claim.
Multi-function installation tester with earth and RCD testing

MODEL

KEW

KEW

Describe your application and constraints in a request.
What is recorded about the regulatory status and the life of this part.
No compliance statement is recorded for this part. Ask the supplier for the RoHS and REACH declarations and, for a radio product, the EU declaration of conformity.
| Current status | Status not stated |
| Key dates | Not recorded |
| Manufacturer part number | KEW_6010B |
| Manufacturer | Kyoritsu |
Installation and Earth Testers
Single and three-phase ZigBee electrical metering module
Price on request
Installation and Earth Testers
Multifunction installation tester with loop and RCD testing · Ref. : KEW_6016
Price on request
Installation and Earth Testers
Advanced earth resistance and soil resistivity tester · Ref. : UT572
Price on request
Installation and Earth Testers
Residual current device (RCD) tester with trip time measurement · Ref. : KEW_5410
Price on request
With your consent, we use Google cookies to measure the site’s audience and our advertising. Your quote basket and comparison stay in your browser either way. Details