FEED – Long-duration data logger with continuous sampling for oscilloscope

Long-duration data logger with continuous sampling for oscilloscope

Data Loggers

Manufacturer part number
FEED
Manufacturer
Siglent Technologies

Long-duration data logging feature of the four-channel Siglent SDS1000X-E oscilloscopes, continuously recording original sampling points and measurement values.

Product overview

Data logging. Siglent Technologies presents the Datalogging mode of its four-channel SDS1000X-E oscilloscope series, which collects data over long durations, from tens of minutes to several hours, without developing acquisition software.

Sample Logger mode. The Sample Logger continuously samples and stores measurement points, suited to long-time observation of low-speed signals. Points are saved in real time to the oscilloscope's internal flash memory or to an external USB drive, since the DDR memory is limited. The log file is recorded in binary format with the *.slg extension, and the data can then be recalled on the oscilloscope or analysed on a computer.

Time stamping. The SDS1000X-E oscilloscopes have no system clock. The time must be set manually or through an NTP network time server, with resynchronisation after power loss or periodically depending on configuration.

Typical uses. This mode suits measurement tests requiring prolonged acquisition, for which a software acquisition chain would otherwise be needed.

Integration. Before logging, configure the scaling, type and activation of the input channels, then calculate the data volume from the sample rate and the desired duration.

See also

Technical questions

Ask a question

No published question about this part yet. Questions are reviewed before publication.

Similar products

All Data Loggers
Request a quote