ME3-CH2O
Electrochemical formaldehyde CH2O sensor, 0 to 10 ppm

- Pressure range
- Standardatmosphere±10%
- Detection range
- 0~10ppm
- Measurement Range
- 0~10ppm
- Storage humidity
- 15﹪~90﹪RH
Electrochemical formaldehyde CH2O sensor, 0 to 10 ppm
Winsen ME3-CH2O electrochemical sensor for measuring formaldehyde (HCHO) concentration over the 0 to 10 ppm range, for industrial and environmental applications.
Overview. The Winsen ME3-CH2O is an electrochemical formaldehyde sensor (CH2O, HCHO) that measures gas concentration from the current produced by electrochemical oxidation of the target gas on the working electrode, this current being proportional to concentration according to Faraday's law.
Main characteristics. Measurement range of 0 to 10 ppm, maximum detectable concentration of 100 ppm, resolution of 0.1 ppm, sensitivity of (11.8 ± 6) µA/ppm, response time T90 of 90 s or less, linear output, repeatability better than 2% of output value and stability better than 2% per month. Zero drift is between -0.03 and 0.03 ppm over -20 to 40 °C. Anticipated service life is 2 years in air.
Electrical and detection principle. Bias voltage of 300 mV and recommended load resistance of 300 Ω. The sensor offers low consumption, high precision, high sensitivity and good anti-interference ability.
Environmental conditions. Standard atmosphere pressure range ±10%. Storage between -20 and 50 °C and between 15 and 90% RH, without condensation. Conditions of temperature and humidity outside these ranges, or prolonged high temperature and high humidity, reduce sensor life.
Typical uses. Formaldehyde measurement in industrial and environmental applications.
Integration. Ageing and calibration are recommended once the sensor has stabilised, according to the programme and operating conditions of the complete device, as transport and storage can alter its performance.
No published question about this part yet. Questions are reviewed before publication.
Every characteristic recorded for ME3-CH2O, with where it comes from. Use the filters to find one quickly.
| Characteristic | Value |
|---|---|
| Pressure range | Standardatmosphere±10% |
| Response time | ≤90S |
| Anticipated using life | 2 years(in air) |
| Bias voltage | 300mV |
| Characteristics | Low consumption,High precision,High sensitivity |
| Detection Gas | CH2O |
| Detection Principle | Electrochemical |
| Detection range | 0~10ppm |
| Load resistance(recommended) | 300&Omega |
| Max detecting concentration | 100ppm |
| Measurement Range | 0~10ppm |
| Model | ME3-CH2O |
| Output Linearity | linear |
| Repeatability | <2﹪output value |
| Resolution and accuracy | (11.8±6)µA/ppm |
| Resolution ratio | 0.1ppm |
| Response Time (T90) | ≤ 90S |
| Sensitivity | (11.8±6)µA/ppm |
| Stability (/month) | <2﹪ |
| Storage humidity | 15﹪~90﹪RH |
| Storage Temperature | -20℃~50℃ |
| Target | CH2O, HCHO, Formaldehyde |
| Zero drift (-20~40℃) | -0.03ppm~0.03ppm |
No specification matches.
Last verification : 2026-11-02
Official documents and the technical data recorded for ME3-CH2O. For a missing document or a specific revision, our experts send it to you.
Winsen ME3-CH2O electrochemical sensor for measuring formaldehyde (HCHO) concentration over the 0 to 10 ppm range, for industrial and environmental applications.
| Parameter | Value |
|---|---|
| Pressure range | Standardatmosphere±10% |
| Response time | ≤90S |
| Anticipated using life | 2 years(in air) |
| Bias voltage | 300mV |
| Characteristics | Low consumption,High precision,High sensitivity |
| Detection Gas | CH2O |
| Detection Principle | Electrochemical |
| Detection range | 0~10ppm |
| Load resistance(recommended) | 300&Omega |
| Max detecting concentration | 100ppm |
| Measurement Range | 0~10ppm |
| Model | ME3-CH2O |
| Output Linearity | linear |
| Repeatability | <2﹪output value |
| Resolution and accuracy | (11.8±6)µA/ppm |
| Resolution ratio | 0.1ppm |
| Response Time (T90) | ≤ 90S |
| Sensitivity | (11.8±6)µA/ppm |
| Stability (/month) | <2﹪ |
| Storage humidity | 15﹪~90﹪RH |
| Storage Temperature | -20℃~50℃ |
| Target | CH2O, HCHO, Formaldehyde |
| Zero drift (-20~40℃) | -0.03ppm~0.03ppm |
Similar parts to answer your needs. Compare the main characteristics. A recorded replacement is a documented relation; a similar product is a similarity only, never a compatibility claim.
Electrochemical formaldehyde CH2O sensor, 0 to 10 ppm

ME2-CH2O-16×15 HCHO Formaldehyde Sensor

Electrochemical HCHO Formaldehyde Sensor

Explosion-proof Formaldehyde Detector Gas Sensor

Describe your application and constraints in a request.
What is recorded about the regulatory status and the life of this part.
No compliance statement is recorded for this part. Ask the supplier for the RoHS and REACH declarations and, for a radio product, the EU declaration of conformity.
| Current status | Status not stated |
| Key dates | Not recorded |
| Manufacturer part number | ME3-CH2O |
| Manufacturer | Winsen |
Formaldehyde Sensors
Digital formaldehyde (HCHO) sensor with I²C interface · Ref. : IAGM1
Price on request
Formaldehyde Sensors
Formaldehyde MEMS sensor, breakout, 0-3 ppm range · Ref. : SEN0563
Price on request
Formaldehyde Sensors
4-in-1 Indoor Air Quality Monitor with Touchscreen, HCHO and TVOC · Ref. : SFD-02
Price on request
Formaldehyde Sensors
Electrochemical formaldehyde CH2O sensor, UART output · Ref. : ZE510-CH2O
Price on request
With your consent, we use Google cookies to measure the site’s audience and our advertising. Your quote basket and comparison stay in your browser either way. Details