MEU-O2
Electrochemical oxygen (O2) sensor 0 to 25% vol

- Detection range
- 0~25% VOL
- Measurement Range
- 0~25% VOL
- Bias
- -600mV
- Detection Gas
- Oxygen(O2)
Electrochemical oxygen (O2) sensor 0 to 25% vol
Fixed potential electrochemical sensor for measuring oxygen concentration over 0 to 25% vol, used for oxygen detection in industry, mines, warehousing and environmental applications.
Overview. The MEu-O2 is a fixed potential electrochemical oxygen O2 sensor manufactured by Winsen. It measures oxygen concentration over a range of 0 to 25% vol, with a maximum range of 30% vol. The current produced is proportional to the oxygen concentration, following Faraday's law.
Main characteristics. The sensor offers a sensitivity of 80 to 130 µA at 20.9% vol., a repeatability better than ±2% of the output value, a zero drift below 1% vol. between -20 and 40 °C and a linear output. The T90 response time is under 15 s and the rated lifespan is 5 years. The detection principle produces no pollution and the sensor provides good long-term stability.
Operating conditions. The sensor operates between -20 and 50 °C, with humidity from 15 to 90% RH without condensation and a pressure of 1 atm ±10%. The bias is -600 mV and the recommended load resistance is 500, 1K or 2K Ω. Outside these ranges, or under prolonged high temperature and high humidity, the sensor lifespan is reduced.
Typical uses. Oxygen measurement in industrial processes, mines, warehouses and environmental applications.
Integration. The sensor has no sleeping mode; one can be set in the program if needed. Calibration after stabilisation is recommended according to the program and operating conditions of the complete device. The MEu-O2 is available on quotation from Atlas Electronics.
No published question about this part yet. Questions are reviewed before publication.
Every characteristic recorded for MEU-O2, with where it comes from. Use the filters to find one quickly.
| Characteristic | Value |
|---|---|
| Bias | -600mV |
| Characteristics | No pollution, long life, high precision, high sensitivity, Excellent repeatability and stability |
| Detection Gas | Oxygen(O2) |
| Detection Principle | Electrochemical |
| Detection range | 0~25% VOL |
| Lifespan | 5 years |
| Load resistance(recommended) | (500/ 1K/2K)&Omega |
| MaxRange | 30% VOL |
| Measurement Range | 0~25% VOL |
| Model | MEu-O2 |
| Output Linearity | Linear |
| Repeatability | <±2% output value |
| Response Time (T90) | <15S |
| Sensitivity(20.9%vol.) | (80~130)uA |
| Target | Oxygen(O2) |
| Working humidity | 15﹪~90﹪RH |
| Working Pressure | 1atm±10% |
| Working temperature | -20℃~50℃ |
| Zero drift (-20~40℃) | <1% VOL |
No specification matches.
Last verification : 2026-11-02
Official documents and the technical data recorded for MEU-O2. For a missing document or a specific revision, our experts send it to you.
Fixed potential electrochemical sensor for measuring oxygen concentration over 0 to 25% vol, used for oxygen detection in industry, mines, warehousing and environmental applications.
| Parameter | Value |
|---|---|
| Bias | -600mV |
| Characteristics | No pollution, long life, high precision, high sensitivity, Excellent repeatability and stability |
| Detection Gas | Oxygen(O2) |
| Detection Principle | Electrochemical |
| Detection range | 0~25% VOL |
| Lifespan | 5 years |
| Load resistance(recommended) | (500/ 1K/2K)&Omega |
| MaxRange | 30% VOL |
| Measurement Range | 0~25% VOL |
| Model | MEu-O2 |
| Output Linearity | Linear |
| Repeatability | <±2% output value |
| Response Time (T90) | <15S |
| Sensitivity(20.9%vol.) | (80~130)uA |
| Target | Oxygen(O2) |
| Working humidity | 15﹪~90﹪RH |
| Working Pressure | 1atm±10% |
| Working temperature | -20℃~50℃ |
| Zero drift (-20~40℃) | <1% VOL |
Similar parts to answer your needs. Compare the main characteristics. A recorded replacement is a documented relation; a similar product is a similarity only, never a compatibility claim.
Electrochemical oxygen (O2) sensor 0 to 25% vol

Oxygen Sensor

Oxygen Gas Sensor

SUP-DO7013 Electrochemical dissolved oxygen sensor

Describe your application and constraints in a request.
What is recorded about the regulatory status and the life of this part.
No compliance statement is recorded for this part. Ask the supplier for the RoHS and REACH declarations and, for a radio product, the EU declaration of conformity.
| Current status | Status not stated |
| Key dates | Not recorded |
| Manufacturer part number | MEU-O2 |
| Manufacturer | Winsen |
Oxygen Sensors
Fluorescence dissolved oxygen sensor with RS-485 and 4-20mA · Ref. : WQ-DO-4A
Price on request
Oxygen Sensors
Drinking Water Dissolved Oxygen Sensor Probe
Price on request
Oxygen Sensors
Wall-Mounted Oxygen Transmitter with RS485 Output
Price on request
Oxygen Sensors
Industrial dissolved oxygen sensor, 0-20 mg/L range · Ref. : DOG-209FA
Price on request
With your consent, we use Google cookies to measure the site’s audience and our advertising. Your quote basket and comparison stay in your browser either way. Details