The Kyoritsu MODEL 8113 is a clamp-on AC/DC current/voltage converter that measures AC and DC currents from 1 mA to 200 A with a voltmeter of 0.1 mV resolution.
Frequency responseDC: - 1kHz
Range — Measuring Ranges — Output Voltage — AccuracyFrequency Response
Kyoritsu MODEL 8113. clamp-on AC/DC current adaptor, a discontinued product that converts an AC or DC current into a voltage readable by a voltmeter.
Main characteristics. AC measuring ranges of 2 A, 20 A and 200 A, with outputs of 100 mV/A, 10 mV/A and 1 mV/A respectively. The 200 A range gives 200 mV at full scale, for measurement from 1 mA to 200 A with a voltmeter of 0.1 mV resolution.
Accuracy and frequency response. Accuracy of ±2 % of reading ±5 mA on the 2 A range, ±1.5 % of reading ±0.04 A on the 20 A range, ±2 % of reading ±0.5 A on the 200 A range up to 150 A, and ±3.5 % of reading from 150 A to 200 A. Frequency response DC to 1 kHz.
Dimensions and aperture. Conductor size Ø 19 mm maximum. Dimensions 180 × 54 × 31 mm for a weight of approximately 170 g.
Power and safety. Powered by two R6 (AA) 1.5 V batteries. Withstand voltage 1000 V AC for 1 minute.
Typical uses. Current measurement on a conductor without breaking the circuit, with the output voltage read on an external voltmeter.
Supplied with. Output cords 7061, two R6 (AA) batteries, carrying case 9055 and instruction manual.
The Kyoritsu MODEL 8113 is a clamp-on AC/DC current/voltage converter that measures AC and DC currents from 1 mA to 200 A with a voltmeter of 0.1 mV resolution.
Frequency responseDC: - 1kHz
Range — Measuring Ranges — Output Voltage — AccuracyFrequency Response
Range — Measuring Ranges — Output Voltage — Accuracy
Frequency Response
Weight
170g approx.
Withstand Voltage
1000V AC for 1 minute
Alternatives and comparable products
Similar parts to answer your needs. Compare the main characteristics. A recorded replacement is a documented relation; a similar product is a similarity only, never a compatibility claim.
Current productKyoritsu
MODEL_8113
AC/DC current clamp adaptor
Frequency response
DC: - 1kHz
Range — Measuring Ranges — Output Voltage — Accuracy
What is recorded about the regulatory status and the life of this part.
Regulatory compliance
No compliance statement is recorded for this part. Ask the supplier for the RoHS and REACH declarations and, for a radio product, the EU declaration of conformity.
No advertising or analytics cookies. A single, strictly necessary cookie is set only if you sign in to your account. Your quote basket and comparison stay in your browser. Details