MODEL 8126 – AC 200 A voltage-output clamp meter

AC 200 A voltage-output clamp meter

Clamp Meters

  • AC 200A
  • Approx. 3m
  • Ø 40mm
  • MINI DIN 6PIN
Manufacturer part number
MODEL_8126
Manufacturer
Kyoritsu
Rated Current
AC 200A
Output voltage
AC 500mV/200A (AC 2.5mV/A)

Kyoritsu MODEL 8126 clamp meter for AC current measurement up to 200 A, with a 2.5 mV/A voltage output. Its Ø40 mm jaw allows measurement on large-section conductors.

Product overview

Kyoritsu MODEL 8126, a clamp meter for AC current measurement up to 200 A. — Main characteristics: the voltage output delivers AC 500 mV for 200 A, that is 2.5 mV/A, with an accuracy of ±0.5%rdg ±0.1 mV at 50/60 Hz and ±1.0%rdg ±0.2 mV from 40 Hz to 1 kHz. The phase shift is within ±1.0° from 45 to 65 Hz. — Connectivity and output: the jaw accepts conductors of Ø40 mm, the cable is approximately 3 m long and the output uses a MINI DIN 6-pin connector, with an output impedance of approximately 5 Ω. — Power and environment: the clamp operates between 0 and 50 °C, with relative humidity of 85 % or less and no condensation. Its dimensions are 128 x 81 x 36 mm for approximately 260 g. — Safety and standards: compliant with IEC 61010-1, IEC 61010-2-032:2002 and IEC 61326, CAT III 600 V, pollution degree 2. — Typical uses: AC current measurement on electrical installations, with a voltage output usable by a recorder or data acquisition system. — Accessories: carrying case 9095, instruction manual and cable markers supplied; extension cable 7185 and banana adapter plugs 7146 optional. Available on quotation from Atlas Electronics.

See all specifications

Highlights

  • Output voltage: AC 500mV/200A (AC 2.5mV/A)
  • Cable Length: Approx. 3m
  • Dimension: 128(L) x 81(W) x 36(D) mm
  • Operating temperature and humidity range: 0 to 50˚C, relative humidity 85% or less (no condensation)
  • Output Connector:: MINI DIN 6PIN
  • Output Impedance: Approx. 5&Omega

Good to know before integrating

  • Thermal: rated operating range 0 to 50˚C, relative humidity 85% or less (no condensation) — check it against the installation environment.

See also