Winsen ME3-CL2 electrochemical sensor for detecting chlorine gas (Cl2) in chemical and pharmaceutical production and environmental protection applications.
Temperature range-20℃~50℃
Pressure rangeStandard atmospheric pressure ±10%
Humidity range15%~90% RH No condensation
Detection range0-10ppm, Max 100ppm
Max Range250ppm
Bias voltage0mV
Product overview
Overview. The Winsen ME3-CL2 gas sensor is an electrochemical cell-type sensor that detects chlorine (Cl2). The current produced by the electrochemical oxidation of the target gas on the working electrode is proportional to its concentration, following Faraday's law.
Main characteristics. Detection range of 0 to 10 ppm, with a maximum of 100 ppm and a stated maximum range of 250 ppm. Resolution of 0.1 ppm, sensitivity of (1±0.25) µA/ppm, linear output, repeatability below 2% of the output value and stability below 2% per month. Zero drift of 0.2 ppm or less between -20 and 40 °C.
Response and lifespan. Response time T90 of 60 s or less. Anticipated service life of 2 years.
Operating conditions. Temperature range of -20 °C to 50 °C and humidity of 15% to 90% RH without condensation. Standard atmospheric pressure ±10%. Recommended load resistance of 10 Ω. Dimensions φ20 × 16.4 mm.
Typical uses. Chlorine monitoring in chemical and pharmaceutical production processes and in environmental protection applications, according to the characteristics stated by the manufacturer.
Integration. The manufacturer specifies that the sensor has no sleep mode, which can be set in software if needed, and recommends ageing and calibration once the sensor has stabilised, according to the device and its use.
Winsen ME3-CL2 electrochemical sensor for detecting chlorine gas (Cl2) in chemical and pharmaceutical production and environmental protection applications.
Temperature range-20℃~50℃
Pressure rangeStandard atmospheric pressure ±10%
Humidity range15%~90% RH No condensation
Detection range0-10ppm, Max 100ppm
Max Range250ppm
2. Datasheet extract (recorded values)
Parameter
Value
Pressure range
Standard atmospheric pressure ±10%
Humidity range
15%~90% RH No condensation
Bias voltage
0mV
Characteristics
Detection of chlorine gas in chemical, pharmaceutical production and environmental protection fields
Detection Gas
Chlorine Gas (Cl2)
Detection Principle
Electrochemical
Detection range
0-10ppm, Max 100ppm
Life span
2 years
Load Resistance
10Ω (recommended)
Max Range
250ppm
Model
ME3-CL2
Output Linearity
Linear
Repeatability
<2% Output value
Resolution
0.1ppm
Response Time (T90)
≤ 60S
Sensitivity
(1±0.25) µA/ppm
Size
φ20×16.4
Stability (/month)
<2%
Target
CL2
Temperature range
-20℃~50℃
Working Conditions
Resolution ratio 0.1ppm; Response time<60s; Anticipated using life 2 years
Zero drift (-20~40℃)
≤ 0.2ppm
Applications and use cases
Architectures in which ME3-CL2 can play a role, derived from its category. This is not a compatibility statement: check the specifications against your requirements.
Offices and classrooms get stuffy without anyone measuring why. Indoor air quality monitoring places CO2, VOC and particle sensors in the rooms and reports their values through a gateway, so ventilation can be adjusted to real occupancy
To check
Parameters to measure (CO2 alone or CO2, VOC and particles)
Battery-powered sensors or mains-powered sensors with display
Integration with the ventilation control or building management system
Associated system blocks
Air sensorsGateway
Need a complete solution?
Send a request for ME3-CL2 and the other parts of your project.
Similar parts to answer your needs. Compare the main characteristics. A recorded replacement is a documented relation; a similar product is a similarity only, never a compatibility claim.
Current productWinsen
ME3-CL2
Electrochemical chlorine gas (Cl2) sensor 0-10 ppm
What is recorded about the regulatory status and the life of this part.
Regulatory compliance
No compliance statement is recorded for this part. Ask the supplier for the RoHS and REACH declarations and, for a radio product, the EU declaration of conformity.