Electrochemical hydrogen (H2) gas sensor for industrial and environmental protection applications, with a measuring range of 0 to 1000 ppm.
Temperature range-20℃~50℃
Pressure rangeStandard atmospheric pressure±10%
Humidity range15%~90% RH non condensation
Detection range0-1000ppm, Max 2000ppm
Max Range2000ppm
Bias voltage0mV
Product overview
Winsen ME4-H2. The Winsen ME4-H2 hydrogen gas sensor is a constant potential electrolysis sensor designed for hydrogen (H2) detection in industrial and environmental protection applications. The current generated is directly proportional to the hydrogen concentration, which allows the concentration of the target gas to be determined.
Main characteristics. Measuring range 0 to 1000 ppm (maximum 2000 ppm), resolution 1 ppm, sensitivity (10.8 ± 5.2) nA/ppm, linear output, repeatability below 2% of output value and stability below 2% per month. Zero drift at most 10 ppm between -20 and 40 °C. Response time T90 at most 30 s. Anticipated service life 2 years. Low power consumption, high precision and sensitivity, wide linear range and good anti-interference ability.
Operating conditions. Temperature range -20 °C to 50 °C, standard atmospheric pressure ±10%, relative humidity 15% to 90% non-condensing. Recommended load resistance 10 Ω. Dimensions φ32 × 14.5 mm.
Typical uses. Hydrogen monitoring in industrial processes and environmental protection. Prolonged exposure to high temperature or high humidity may reduce the sensor's service life.
Integration. The sensor has no sleeping mode; one can be set in the program if needed. Calibration after stabilisation is recommended according to the complete device and its operating conditions. Available on quotation from Atlas Electronics.
Electrochemical hydrogen (H2) gas sensor for industrial and environmental protection applications, with a measuring range of 0 to 1000 ppm.
Temperature range-20℃~50℃
Pressure rangeStandard atmospheric pressure±10%
Humidity range15%~90% RH non condensation
Detection range0-1000ppm, Max 2000ppm
Max Range2000ppm
2. Datasheet extract (recorded values)
Parameter
Value
Pressure range
Standard atmospheric pressure±10%
Humidity range
15%~90% RH non condensation
Bias voltage
0mV
Characteristics
Detection of hydrogen in industrial and environmental protection fields
Detection Gas
Hydrogen (H2)
Detection Principle
Electrochemical
Detection range
0-1000ppm, Max 2000ppm
Life span
2 years
Load Resistance
10Ω (recommended)
Max Range
2000ppm
Model
ME4-H2
Output Linearity
Linear
Repeatability
<2% Output value
Resolution
1ppm
Response Time (T90)
≤ 30S
Sensitivity
(10.8±5.2) nA/ppm
Size
φ32×14.5
Stability (/month)
<2%
Target
H2
Temperature range
-20℃~50℃
Working Conditions
Resolution ratio 1ppm; Response time≤30s; Anticipated using life 2 years
Zero drift (-20~40℃)
≤ 10ppm
Applications and use cases
Architectures in which ME4-H2 can play a role, derived from its category. This is not a compatibility statement: check the specifications against your requirements.
Offices and classrooms get stuffy without anyone measuring why. Indoor air quality monitoring places CO2, VOC and particle sensors in the rooms and reports their values through a gateway, so ventilation can be adjusted to real occupancy
To check
Parameters to measure (CO2 alone or CO2, VOC and particles)
Battery-powered sensors or mains-powered sensors with display
Integration with the ventilation control or building management system
Associated system blocks
Air sensorsGateway
Need a complete solution?
Send a request for ME4-H2 and the other parts of your project.
Similar parts to answer your needs. Compare the main characteristics. A recorded replacement is a documented relation; a similar product is a similarity only, never a compatibility claim.
What is recorded about the regulatory status and the life of this part.
Regulatory compliance
No compliance statement is recorded for this part. Ask the supplier for the RoHS and REACH declarations and, for a radio product, the EU declaration of conformity.
With your consent, we use Google cookies to measure the site’s audience and our advertising. Your quote basket and comparison stay in your browser either way. Details