KOG-206
Optical Dissolved Oxygen Sensor, 4-20 mA and RS485 Output

- Pressure range
- ≤0.3Mpa
- Measurement Range
- DO:0-20 mg/L,0-20 ppm; Temperature:0-45℃
- Storage Temperature
- -15-65℃
- Communication Protocol
- RS485(customize)
Optical Dissolved Oxygen Sensor, 4-20 mA and RS485 Output
The Apure KOG-206 optical dissolved oxygen sensor measures dissolved oxygen in water by luminescence, with a matched transmitter providing analogue outputs and RS485 communication.
Overview. The Apure KOG-206 is an optical dissolved oxygen sensor using an oxygen-sensitive luminescent layer, supplied with a transmitter. It measures dissolved oxygen from 0 to 20 mg/L (0-20 ppm) and temperature from 0 to 45 °C.
Accuracy and measurement. Stated accuracy is ±3% of the measured value for dissolved oxygen and ±0.5 °C for temperature. Calibration is performed in air or on a sample, according to the procedure of the supplied controller.
Connectivity and outputs. The transmitter provides 2-way 4-20 mA outputs and an RS-485 interface with a customizable communication protocol. The standard cable length is 10 m and may be extended up to 100 m.
Power and environment. Power supply of 100-240 VAC (50/60 Hz) or 9-36 VDC. The sensor is rated IP68 / NEMA 6P and the transmitter IP65 / NEMA 4X. Body in SUS316L and PVC for fresh water, titanium alloy for marine use. Operating pressure up to 0.3 MPa, storage temperature from -15 to 65 °C.
Dimensions. Sensor 55 mm in diameter and 342 mm long, transmitter 142 x 142 x 162 mm (L x W x H).
Integration. Dissolved oxygen measurement in fresh or marine water, to be integrated into a control system through the analogue outputs or the RS485 link.
No published question about this part yet. Questions are reviewed before publication.
Every characteristic recorded for KOG-206, with where it comes from. Use the filters to find one quickly.
| Characteristic | Value |
|---|---|
| Dimensions | Sensor:Dia55mm*L 342mm; Transmitter:142*142*162mm(L*W*H) |
| Pressure range | ≤0.3Mpa |
| Cable Length | Standard:10m, the maximum may be extended to 100m |
| Calibration | Air or sample calibration according to the supplied controller procedure |
| Communication Protocol | RS485(customize) |
| Main Material | Body:SUS316L + PVC(fresh water),Titanium alloy(marine); O-ring:fluorine rubber; Cable:PVC |
| Measurement Accuracy | DO:±3% of the measured value; Temperature:±0.5℃ |
| Measurement Range | DO:0-20 mg/L,0-20 ppm; Temperature:0-45℃ |
| Measuring temperature | 0-45℃ |
| Output | 2-way 4-20mA |
| Power supply | AC:100-240VAC (50/60HZ); DC:9~36VDC |
| Protective rate | Sensor:IP68/NEMA6P; Transmitter:IP65/ NEMA4X |
| Storage Temperature | -15-65℃ |
| Weight | Sensor:1.85KG; Transmitter:1.35KG |
No specification matches.
Last verification : 2026-11-06
Official documents and the technical data recorded for KOG-206. For a missing document or a specific revision, our experts send it to you.
The Apure KOG-206 optical dissolved oxygen sensor measures dissolved oxygen in water by luminescence, with a matched transmitter providing analogue outputs and RS485 communication.
| Parameter | Value |
|---|---|
| Dimensions | Sensor:Dia55mm*L 342mm; Transmitter:142*142*162mm(L*W*H) |
| Pressure range | ≤0.3Mpa |
| Cable Length | Standard:10m, the maximum may be extended to 100m |
| Calibration | Air or sample calibration according to the supplied controller procedure |
| Communication Protocol | RS485(customize) |
| Main Material | Body:SUS316L + PVC(fresh water),Titanium alloy(marine); O-ring:fluorine rubber; Cable:PVC |
| Measurement Accuracy | DO:±3% of the measured value; Temperature:±0.5℃ |
| Measurement Range | DO:0-20 mg/L,0-20 ppm; Temperature:0-45℃ |
| Measuring temperature | 0-45℃ |
| Output | 2-way 4-20mA |
| Power supply | AC:100-240VAC (50/60HZ); DC:9~36VDC |
| Protective rate | Sensor:IP68/NEMA6P; Transmitter:IP65/ NEMA4X |
| Storage Temperature | -15-65℃ |
| Weight | Sensor:1.85KG; Transmitter:1.35KG |
Answers to the questions that the recorded characteristics of KOG-206 can answer. A question is not asked when nothing is recorded to answer it.
Yes — recorded under "Communication Protocol": RS485(customize).
Similar parts to answer your needs. Compare the main characteristics. A recorded replacement is a documented relation; a similar product is a similarity only, never a compatibility claim.
Optical Dissolved Oxygen Sensor, 4-20 mA and RS485 Output

SUP-DO7013 Electrochemical dissolved oxygen sensor

SUP-DO7017 Optical dissolved oxygen sensor

SUP-DO7018 Optical dissolved oxygen sensor

Describe your application and constraints in a request.
What is recorded about the regulatory status and the life of this part.
No compliance statement is recorded for this part. Ask the supplier for the RoHS and REACH declarations and, for a radio product, the EU declaration of conformity.
| Current status | Status not stated |
| Key dates | Not recorded |
| Manufacturer part number | KOG-206 |
| Manufacturer | Apure |
Oxygen Sensors
Fluorescence dissolved oxygen sensor with RS-485 and 4-20mA · Ref. : WQ-DO-4A
Price on request
Oxygen Sensors
Drinking Water Dissolved Oxygen Sensor Probe
Price on request
Oxygen Sensors
Wall-Mounted Oxygen Transmitter with RS485 Output
Price on request
Oxygen Sensors
Industrial dissolved oxygen sensor, 0-20 mg/L range · Ref. : DOG-209FA
Price on request
No advertising or analytics cookies. A single, strictly necessary cookie is set only if you sign in to your account. Your quote basket and comparison stay in your browser. Details